| Feature: |
| Capsu-Photohelic® Switch/Gages function as versatile, highly repeatable pressure switches combined with a precise pressure gage employing the time proven Magnehelic® gage design. The Capsu-Photohelic® gage employs an encapsulated sensing element for use with both liquids and gases at pressures to 500 psig (34 bar). Optional cast brass case is available for water or water based liquids. Two phototransistor actuated, DPDT relays are included for low/high limit control. Easy to adjust setpoint indicators are controlled by knobs located on the gage face. Individual setpoint deadband is one pointer width — less than 1% of full scale. Set-points can be interlocked to provide variable deadband — ideal for control of pumps, etc. |
| Specifications: |
| Gauge Specifications |
| Service | Compatible gases and liquids. Brass case option required for water based liquids. |
| Wetted Materials | Chat with Agent |
| Accuracy | ±3% of full scale at 70°F (21.1°C). ±2% on 43000S models; ±4% on 43215, 43220 and 43500. |
| Pressure Limits | -20 Hg to 500 psig (-0.677 to 34.5 bar). |
| Temperature Limits | 20 to 120°F. (-6.67 to 48.9°C) Low temperature option available. |
| Process Connections | 1/4˝ female NPT. |
| Size | 4˝ (101.6 mm) dial face, 5˝ (127 mm) O.D. x 9-3/16˝ (233.36 mm). |
| Weight | 5 lb, 8 oz (2.49 kg). Brass 11 lb, 2 oz (5.05 kg). |
|
|
|
| Switch Specifications |
| Switch Type | Each setpoint has 2 Form C relays (DPDT). |
| Repeatability | ±1% of full scale. |
| Electrical Rating | 10A @ 120 VAC, 6A @ 240 VAC, 60 Hz res. 10A @ 28 VDC. |
| Electrical Connections | Screw Terminals. |
| Power Requirements | 120 VAC, 50/60 Hz; 240 VAC & 24 VAC power optional. |
| Mounting Orientation | Diaphragm in vertical position. Consult factory for other position orientations. |
| Set Point Adjustment | Adjustable knobs on face. |