PDFPrintE-mail
Dwyer-Spirit-Filled Glass Thermometers-GTEurotron-Low Cost IR thermometer- Micro

Eurotron-Low Cost IR thermometer

( Eurotron )

General:

Emissivity

The infrared energy emitted by a surface differs according to its composition and physical condition. The parameter, used to characterize the object surface, is called emissivity. The preset value 0.95 is valid on most common applications with few exception like shiny metals (e.g. chimney can be measured by painting a small area or sticking a small label). Emissivity can also be adjusted from 0.2 to 1.

Laser Sighting

US Ultra Sharp All MicroRay models are equipped with a Class II Laser for aiming purposes. The offset laser sighting will allow you to identify the approx. center of the target being measured. The Ultra Sharp definition of laser makes it easier to read outdoors as well.

9V Battery

Easily replaceable standard 9V battery is provided with the unit. A kit is also available for rechargeable battery and external charger.

Tripod mounting

Standard photographic tripod thread is available for accurate and long term measurements.

Built-In Contact Probe

The MicroRay HVAC+ is equipped with a connector for contact measurement. The available temperature probes include a Contact, Immersion, Penetration, and Ambient Air. One compact probe can actually be stored inside the unit for easy access. Longer probes are available if needed and provided with a wire connection.

Built-in T+RH% sensor

The MicroRay HVAC++ is equipped with a sensor for ambient Temperature and Relative

Humidity measurements. The internal Sensor allows to check the comfort index in offices and living rooms.

Ranges: -10 to 50°C - 10 to 95%rH

Accuracies: ±0.5°C - ±3%rH

Memory

The MicroRay HVAC++ model can store up to 64 test with date and time using the internal real time clock. Data logging can be manual or automatic with settable sampling time.

Specifications:

Spectral Response

8-14 m m

Display Resolution

HVAC/HVAC+: 0.1°C up to 199.9°C. 1°C otherwise.

HVAC/HVAC+: 0.2°C from -30 to 0°C

HVAC++: 0.1°C

Response Time

500 ms @ 95% response

Emissivity

adjustable from 0.20 to 1.00 (preset on 0.95)

Auto Power Off

Automatic if no key is pressed for more than 30 sec.

Accuracy IR Channel

HVAC++: from 0°C to 70°C : ±1°C From 70°C to f.s. : ±1.5% of the reading or ±1.5°C

HVAC: from 0°C to F.S.: ±1.5% of the reading or ±1.5°C

All models: from -30°C to 0°C : ±(1°C + 0.07°C/°C)

Repeatability IR Channel

±0.5% of the reading or ±0.7°C from 0°C

External Channel

Resistance high accuracy platinum Pt1000

External Channel Range

from -30°C to 500°C

External Channel Accuracy:

0.2% of reading or 0.2°C

Power Supply

9V alkaline or optional rechargeable battery kit

Battery Life

40 h (back light and laser Off) - 20 h (On)

Operative Temperature

0 to +50°C 10-95% RH non condensing up to 30°C

Storage Temperature

-20°C to +60°C without Battery

Tripod mounting

6.35 mm ( 1/4” ) 20 UNC Threading

Dimensions – Weight

140x90x38 mm - 165g Nett