PDFPrintE-mail
Smart Loop Calibrator-LC 21

Thermocouple Calibrator

( Dwyer )

General

Dual function CA10 Thermocouple Calibrator can be used to source Type K thermocouple output values or measure Type K thermocouples. Quickly select output value using the rotary switch with coarse and fine adjustments. Model CA10 accepts up to two Type K thermocouples and features data hold and max functions. Measurements are displayed on a 3-1/2 digit backlit display. Units include a 4˝ (100 m) Type K bead wire temperature probe, calibration cable, 9V battery and instruction manual.

 

 

Specifications:

Measurement Range

-40 to 1400°F (-40 to 760°C), selectable.

Thermocouple Input

Type K.

Input Protection

24 VDC or 24 VAC rms max input voltage on any combination of input pins.

Accuracy

±0.1% rdg. + 2°F on -40 to 1400°F; ±0.1% rdg. +1°C on -40 to 760°C, over operating temperature of 64 to 82°F (18 to 28°C).

Output Range:

-328 to 2192°F (-210 to 1200°C).

Display

0.8˝ (20 mm) height, 3-1/2- digit LCD with switchable back light.

Resolution

Selectable 1° or 0.1°.

Reading Rate

2.5 times/second.

Ambient Operating Temperature

32 to 122°F (0 to 50°C) max 80% RH.

Storage Temperature

-4 to 140°F (-20 to 60°C) max 70% RH.

Thermocouple Connection

Standard female mini-connector

Power Requirements

9V Alkaline (included).

Battery Life

200 hours typical

Temperature Coefficient

0.1 times the applicable accuracy specification per °F from 32 to 64°F and 82 to 122°F (0 to 18°C and 28 to 50°C).

Housing Material

ABS plastic.

Weight

15.5 oz (440 g).